• Title of article

    Pseudo-no-reflow phenomenon in ostial saphenous vein graft intervention using FilterWire EX protection

  • Author/Authors

    Chi Hang Lee، نويسنده , , Adrian Low، نويسنده , , Huay Cheem Tan، نويسنده , , Yean Teng Lim، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    2
  • From page
    233
  • To page
    234
  • Abstract
    Distal protection using FilterWire EX has been shown to be effective in reducing the incidence of debris embolization and no-reflow phenomenon in saphenous vein graft (SVG) intervention. We present a case of mechanical obstruction of antegrade flow by atherothrombotic debris collected in the FilterWire EX after stent implantation of the ostial SVG lesion, masqueraded as no-reflow phenomenon. The condition was promptly reverted by thrombosuction using an Export Aspiration Catheter. A heavy debris burden was recovered by the thrombosuction and there was no evidence of myonecrosis after the procedure. We emphasize that different distal protection device can complement each other in special situations. A brief overview of the various approaches in preventing distal embolization in SVG intervention was also given.
  • Keywords
    Saphenous vein graft , Filter wire , distal protection
  • Journal title
    International Journal of Cardiology
  • Serial Year
    2005
  • Journal title
    International Journal of Cardiology
  • Record number