Title of article :
Angioplasty through a stent side door
Author/Authors :
A. Abhyankar، نويسنده , , Meganathan Thirumal and Pratibha L. Gai ، نويسنده , , B.P. Bailey، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Abstract :
Angioplasty was performed by entering a Gianturco-Roubin stent from the side through the coils of a previously-implanted hanging stent in an ostial lesion in a saphenous vein graft. No complications and no visible damage occurred to the stent. Although this method carries a possible risk of entrapment or damage to the stent, it is feasible when other options have failed.
Keywords :
Gianturco-Roubin stent , Ostial , restenosis
Journal title :
International Journal of Cardiology
Journal title :
International Journal of Cardiology