Title of article
Graded glass coatings for Co-Cr implant alloys
Author/Authors
S. Fujino، نويسنده , , H. Tokunaga، نويسنده , , S. Hata، نويسنده , , E.SAIZ، نويسنده , , A. P. Tomsia، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2005
Pages
5
From page
2499
To page
2503
Abstract
A graded glass coating for Vitallium R, a Co-Cr alloy, has been prepared using a simple
enameling technique. The composition of the glasses has been tailored to match the
thermal expansion of the alloys. The optimum glass composition and firing conditions
(temperature and time) needed to fabricate homogeneous coatings with good adhesion to
the alloy were determined. The final coating thickness ranged between 25 and 60 μm.
Coatings fired under optimum conditions do not delaminate during indentation tests of
adhesion. Excellent adhesion to the alloy has been achieved through the formation of 100
nm thick interfacial chromium-oxide (CrOx) layers. The graded glass (consisting of BIG and
6P50 layers) can be successfully coated to a Co-Cr alloy, and forms hydroxyapatite (HA) on
the coating surface when immersed in a simulated body fluid (SBF) for 30 days.
C 2005 Springer Science + Business Media, Inc.