Title of article
Development of the mini-ionization chamber for high-dose real-time monitoring inside a gamma irradiation facility
Author/Authors
M.M.، Hamada, نويسنده , , Ad.A.، Rodrigues, نويسنده , , J.M.، Vieira, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-1097
From page
1098
To page
0
Abstract
A cylindrical ionization chamber of 0.9 cm/sup 3/ has been developed for highdoses real-time monitoring during the sample irradiation at a static position in a /sup 60/Co gamma industrial plant, with about 25.9 PBq (700 kCi). Nitrogen gas at pressure of 1 bar was utilized to fill the ionization chamber for which an appropriate configuration was determined, to be used as a detector for high-dose measurements. The first detector tests were carried out using a /sup 60/Co Gammacell with about 0.26 PBq (7 kCi) and dose rates of 0.6, 1.8, and 2.6 kGy/h. These dose rates were reached with lead absorbers. A good linearity of the detector was found between dose and accumulated charge, independently of the different dose rates. The developed ionization chamber was suitable to be used as a realtime dosimeter for dose rates between 0.6 to 2.6 kGy/h at a /sup 60/Co gammacell, independently of the spectrum degradation caused by absorbers between 50%-90%.
Keywords
rectangular waveguide (RWG) , waveguide transition , millimeter wave , low-temperature co-fired ceramic (LTCC) , Laminated waveguide
Journal title
IEEE Transactions on Nuclear Science
Serial Year
2003
Journal title
IEEE Transactions on Nuclear Science
Record number
86466
Link To Document