Title of article :
Comparison of independent optical frequency measurements using a portable iodine-stabilized Nd:YAG laser
Author/Authors :
Ye، Jun نويسنده , , Hong، Feng-Lei نويسنده , , H.، Matsumoto, نويسنده , , J.L.، Hall, نويسنده , , J.، Ishikawa, نويسنده , , K.، Sugiyama, نويسنده , , A.، Onae, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-23
From page :
24
To page :
0
Abstract :
A comparison of two independent absolute optical frequency measurements has been carried out between the National Metrology Institute of Japan/National Institute of Advanced Industrial Science and Technology (NMIJ/AIST), Tsukuba, Japan and the JILA (formerly the Joint Institute for Laboratory Astrophysics), Boulder, CO, using a portable iodine-stabilized Nd:YAG laser. The agreement between the two absolute measurements is better than the measurement uncertainty of 6.7*10^-13 that can be attributed to the reproducibility limitations of the portable laser. This comparison is used to confirm the measured absolute frequency of an iodine-stabilized Nd:YAG laser at NMIJ/AIST (Y3), which is reported to the Consultative Committee for Length (CCL) for the determination of the absolute frequency value of iodine-stabilized Nd:YAG lasers.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91455