Title of article :
Optical frequency standards based on the /sup 199/Hg/sup +/ ion
Author/Authors :
U.، Tanaka, نويسنده , , J.C.، Bergquist, نويسنده , , S.، Bize, نويسنده , , S.A.، Diddams, نويسنده , , R.E.، Drullinger, نويسنده , , L.، Hollberg, نويسنده , , W.M.، Itano, نويسنده , , C.E.، Tanner, نويسنده , , D.J.، Wineland, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-244
From page :
245
To page :
0
Abstract :
We report on work directed toward the systematic evaluation of an optical frequency standard based on the /sup 2/S/sub 1/2/-/sup 2/D/sub 5/2/ transition of a single, laser-cooled, trapped /sup 199/Hg/sup +/ ion, whose resonance frequency is 1.06*10^15 Hz. For the purpose of the evaluation, a second /sup 199/Hg/sup +/ standard has been constructed. In the cooling-laser system built for the second standard, an injection-locking scheme has been applied to a CW Ti-sapphire laser. We also report optical frequency measurements of the clock transition performed over the past 21 months with the first standard. During this term, the variation of the clock transition frequency is found to be less than +-1*10^-14.