• Title of article

    Development of a cesium atomic fountain frequency standard

  • Author/Authors

    Park، Sang-Eon نويسنده , , Lee، Ho Seong نويسنده , , Kwon، Taeg Yong نويسنده , , Yang، Sung Hoon نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -262
  • From page
    263
  • To page
    0
  • Abstract
    We report on the development of a cesium atomic fountain frequency standard at the Korea Research Institute of Standards and Science (KRISS). Cesium atoms were launched upwards by using an optical moving molasses method after they were cooled in a magnetooptical trap. The temperature of launched atoms was about 2.5 (mu)K. From our cesium atomic fountain, we were able to observe Ramsey fringes of 1-Hz linewidth on the 9.2-GHz clock transition of cesium atoms.
  • Keywords
    leukemia
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Serial Year
    2003
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Record number