Title of article :
Frequency band requirement for a precision RMS measurement system based on successive extraction of the means of absolute values
Author/Authors :
U.، Pogliano, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-336
From page :
337
To page :
0
Abstract :
This paper investigates the requirements for the gain and the frequency band of the components used in the implementation of a precision root mean square measurement system based on the method of successive extraction of the means of the absolute values. The analysis shows the feasibility of a system using this method at a level of accuracy of one part in 10^6 in the frequency range from 10 to 100 Hz.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91477