Title of article :
Calibration of electronic capacitance and dissipation factor bridges
Author/Authors :
G.، Ramm, نويسنده , , H.، Moser, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-395
From page :
396
To page :
0
Abstract :
A new calibration procedure for electronic capacitance and dissipation factor bridges is presented. These bridges measure the ratio of their input currents. For the calibration, two sinusoidal alternating currents are synthesized with the aid of a digital, programmable two-channel AC voltage source and two AC resistors. The complex ratio of these synthesized input currents is calculated from an AC voltage ratio determined by a sampling method, together with the impedance ratio of the AC resistors. At currents of a few mA, their complex ratio in the frequency range up to 100 Hz can be measured with uncertainties (k=2) of approximately 1 . 10^-6 and 1 (mu)rad, respectively.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91491
Link To Document :
بازگشت