Title of article :
Projecting Zener DC reference performance between calibrations
Author/Authors :
C.A.، Hamilton, نويسنده , , L.W.، Tarr, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
Two simple empirical models are proposed to predict Zener reference standard output voltage and uncertainty based on just one or two calibration points. The models are derived from a study of 140 Zener standards. The results yield a substantial reduction in uncertainty relative to the manufacturerʹs specification and illustrate the dominant influence of seasonal effects on projection uncertainty.
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Journal title :