Title of article :
AC-DC transfer comparator for the calibration of thermal voltage converters against Josephson alternating voltage standards
Author/Authors :
H.، Sasaki, نويسنده , , I.، Budovsky, نويسنده , , P.، Coogan, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
This paper describes a new measurement system for the comparison of alternating voltage standards based on the Josephson effect with thermal voltage converters. The target is to achieve an uncertainty of 0.1 (mu) V/V at frequencies up to 10 kHz. The present work achieves an uncertainty of 0.2 (mu) V/V up to 2 kHz.
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT