Title of article :
AC Josephson voltage standard error measurements and analysis
Author/Authors :
C.J.، Burroughs, نويسنده , , S.P.، Benz, نويسنده , , P.D.، Dresselhaus, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-541
From page :
542
To page :
0
Abstract :
The Josephson arbitrary waveform synthesizer can be used as a precision voltage source for both AC and DC signals. Recent improvements in circuit designs have resulted in output voltages greater than 100 mV rms for AC metrology measurements in the frequency range from 1 to 50 kHz. In this paper, we present measurements of several significant error sources that must be considered in order to create an intrinsic AC voltage standard at these frequencies. These results and the corresponding theoretical analysis allow us to move toward creating a practical AC Josephson voltage standard.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91526
Link To Document :
بازگشت