Title of article :
Direct determination of capacitance standards using a quadrature bridge and a pair of quantized Hall resistors
Author/Authors :
B.M.، Wood, نويسنده , , A.D.، Inglis, نويسنده , , M.، Cote, نويسنده , , R.B.، Young, نويسنده , , M.D.، Early, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-558
From page :
559
To page :
0
Abstract :
We have used a quadrature bridge to compare directly two quantum Hall resistors with two 10-nF capacitors at 1233 Hz. This approach eliminates uncertainties due to thermal and structural instabilities and to AC/DC differences of the wire resistors normally used in an intermediate step. We have measured the capacitative interaction between the devices, and have developed a shielded header to suppress this effect. The value for the National Research Council primary standard 10-pF capacitor obtained by this method agrees with the expected value to within (0.1 +- 3.5)aF, or (0.01 +- 0.35) * 10^-6.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91530
Link To Document :
بازگشت