Title of article :
1:30000 cryogenic current comparator with optimum SQUID readout
Author/Authors :
G.، Rietveld, نويسنده , , J.، Sese, نويسنده , , E.، Bartolome, نويسنده , , A.، Camon, نويسنده , , J.، Flokstra, نويسنده , , C.، Rillo, نويسنده , , P.، de la Court, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-620
From page :
621
To page :
0
Abstract :
We have developed a 1:30000 winding ratio cryogenic current comparator (CCC) for measurement of small currents, e.g., generated by a single electron transport device. A dedicated low-noise SQUID is used for the readout, which allows reaching a sensitivity close to ideal, that is without loss of sensitivity in the flux transformer coupling the CCC and the SQUID. The CCC-SQUID equivalent current input noise using a 1:30 000 winding ratio is 2.1 fA/Hz/sup 1/2/, measured down to 0.1 Hz. Room temperature impedance measurements on CCCs appear to be a useful method for the determination of the screening by shields surrounding the CCC as well as for the detection of internal CCC resonances.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91543
Link To Document :
بازگشت