Title of article :
Flotation measurements to reveal the reason for the discrepancy in the molar volume of silicon
Author/Authors :
H.، Bettin, نويسنده , , H.، Toth, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
To reveal the reason for the discrepancy of relatively 3 . 10^-6 in the molar volume of silicon found in crystals of different origin, the densities of crystals grown under different conditions were compared using a flotation method. The method and the apparatus used at Physikalisch-Technische Bundesanstalt, Braunschweig, Germany, and the evaluation of data using Lagrange multipliers are described. The obtained results are reported and the contributions to the standard uncertainty of the silicon density (rho)/sub si/ of about 1.0 . 10^-7 (rho)/sub si/ are discussed.
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT