• Title of article

    A nonlinear dynamic S/H-ADC device model based on a modified Volterra series: identification procedure and commercial CAD tool implementation

  • Author/Authors

    P.A.، Traverso, نويسنده , , D.، Mirri, نويسنده , , G.، Pasini, نويسنده , , F.، Filicori, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -1128
  • From page
    1129
  • To page
    0
  • Abstract
    A nonlinear, dynamic empirical model, based on a Volterra-like approach, was previously proposed by the authors for the time-oriented characterization of sample/hold (S/H) and analog-to-digital conversion (ADC) devices. In this paper, the experimental procedure for model parameter measurement is presented, as well as techniques devoted to the implementation of the model in the framework of the main commercial CAD tools for circuit analysis and design. Examples of simulations, performed both in the time and frequency domain on the model obtained for a commercial device, are proposed, which show the modelʹs capability of pointing out the dynamic nonlinear effects in the S/H-ADC response.
  • Keywords
    leukemia
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Serial Year
    2003
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Record number

    91620