Title of article :
Reflectometer calibration without an open circuit
Author/Authors :
J.، Achkar, نويسنده , , S.A.، Chahine, نويسنده , , B.، Huyart, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-1487
From page :
1488
To page :
0
Abstract :
This paper presents a convenient calibration technique of a six-port reflectometer requiring a set of loads including two standards and avoiding the use of an open circuit. This method has been applied for different reflectometers at three frequencies: 2, 33, and 94 GHz. Reflection coefficient measurement results obtained by this method and other calibration techniques prove its efficiency. Transmission coefficient measurement has also been performed at 94 GHz. Characterizations of some coaxial components by the heterodyne network analyzer using this technique and other calibration methods are presented at frequencies between 0.5 and 10 GHz.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91668
Link To Document :
بازگشت