• Title of article

    A traveling standard for the calibration of data acquisition boards

  • Author/Authors

    A.، Carullo, نويسنده , , M.، Parvis, نويسنده , , A.، Vallan, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    -556
  • From page
    557
  • To page
    0
  • Abstract
    The large use of measurement systems based on data acquisition boards makes the traceability-chain assurance a tricky problem due to the difficulty in consistently calibrating such boards. In this paper, the authors describe a traveling standard which can be used for the calibration of many commercially available acquisition boards. By employing such a traveling standard, the calibration procedure can be remotely exercised by a calibration laboratory through the personal computer which hosts the board that has to be calibrated. In such a way, the calibration results refer to environmental, software, and hardware conditions that exactly match the board-operating conditions. Furthermore, the board unavailability time is drastically reduced, with a consequent economic advantage for the board owner. The traveling standard is based on a microcontroller which is responsible for the communication with the PC that hosts the board and for the boardstimulus generation, and on a digital multimeter, which acts as a reference standard.
  • Keywords
    Power-aware
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Serial Year
    2004
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Record number

    91797