Title of article :
Accuracy of a cryocooler-based programmable Josephson Voltage standard
Author/Authors :
M.، Schubert, نويسنده , , R.، Behr, نويسنده , , T.، May, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A cryocooler-based programmable Josephson array voltage standard (JAVS) for 4-K operation and a very compact microwave source are described. Both components are crucial for future mobile voltage standards. A direct comparison between two programmable JAVS using these components was performed at 1 V in order to test whether these systems are suitable for metrological application at the highest level of precision. The results of the comparison show that the new components are suitable for mobile JAVS with an uncertainty of less than 1 * 10/sup -9/.
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Journal title :