Title of article :
Guidelines for the achievement of true atomic resolution with
noncontact atomic force microscopy
Author/Authors :
Seizo Morita، نويسنده , , Yasuhiro Sugawara، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
We investigated the conditions to achieve true atomic resolution with an atomic force microscope under noncontact mode
NC-AFM.. At first, we derived the equation of vertical resolution as a function of the signal-to-noise ratio and the decay
length of frequency shift by assuming an exponential tip-to-sample distance dependence of frequency shift. Next, by
assuming a single atom probe, we derived the equation of lateral resolution as a function of the vertical resolution and the
tip-to-sample distance, from which we clarified the guidelines to achieve true atomic resolution with NC-AFM. At last, we
made clear the attainable decay length of frequency shift both for the van der Waals potential and the electrostatic
Coulomb.potential. q1999 Elsevier Science B.V. All rights reserved.
Keywords :
Noncontact , resolution , Atomic force microscope , True atomic resolution
Journal title :
Applied Surface Science
Journal title :