• Title of article

    Cation Mass Spectrometer: towards an optimisation of MCsx+ cluster analysis

  • Author/Authors

    T Wirtz، نويسنده , , H.-N Migeon، نويسنده , , H Scherrer، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    5
  • From page
    189
  • To page
    193
  • Abstract
    The Cation Mass Spectrometer (CMS) should lead to high MCsx+ useful yields by allowing to reach an optimum value of the stationary Cs surface concentration, which is known to be a very critical parameter in MCsx+ analysis due to its strong influence on the probability of secondary Cs+ ionisation. For the present work, variations of the Cs concentration were obtained by modifying the impact angle of the primary Cs+ beam and by bombarding the sample simultaneously with Cs+ and Ga+ beams. Results show that these techniques allow to cover a large and continuous range of Cs surface concentrations and to reach an optimisation of the MCsx+ useful yields (enhancements up to a factor 100 compared to Cameca 4f and 6f instruments) mainly due to work function conditions resulting in high Cs+ ionisation probabilities.
  • Keywords
    Useful yield , Cesium concentration , Cs/Ga co-bombardment , Cesium clusters
  • Journal title
    Applied Surface Science
  • Serial Year
    2003
  • Journal title
    Applied Surface Science
  • Record number