Title of article :
Achieving high reproducibility isotope ratios with the
Cameca IMS 1270 in the multicollection mode
Author/Authors :
M. Schuhmacher*، نويسنده , , F. Fernandes، نويسنده , , E. de Chambost، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
Multicollection system allows improvement of measurement reproducibility for isotope ratios on SIMS instruments provided
that the other sources of measurement variability are well controlled. This paper presents analytical protocols developed to
control the mass fractionation in the collection optical system and the variation of electron multiplier yield. Results for both
silicon and oxygen isotope ratios demonstrate that an experimental error as low 0.2 per mil is achievable whenever both isotopes
were measured either with Faraday cup detectors or combining one Faraday cup and one electron multiplier (EM).
# 2004 Elsevier B.V. All rights reserved
Keywords :
Multicollection , isotope ratios , reproducibility , Silicon , Oxygen
Journal title :
Applied Surface Science
Journal title :
Applied Surface Science