A simple, rapid, and sensitive spectrophotometric method for the trace level determination of beryllium based on the
formation of a 1:2 complex with anthralin (1,8-dihydroxyanthrone) as a new reagent is developed. A spectrophotometric method
was used to determine the acidity constant and stepwise proton dissociation of the reagent. The experimental conditions for
determining beryllium including the influences of pH, reagent concentration and time were evaluated and optimized. Under the
optimum experimental conditions, the molar absorptivity of the complex is 0.47 پ~ 104 l mol-1 cm-1 at 545 nm. Calibration graph
was linear in the range of 0.04-1.04 ƒتg ml-1 with a detection limit of 0.012 ƒتg ml-1 and a %RSD of 0.43%, for 5 replicate
determinations at 0.48 ƒتg ml-1 of Be(ƒ§ƒ§). The interferring effect of some cations and anions was also studied. The method was
applied for the determination of beryllium in beryl, silicate rock and alloys. Ethylenediaminetetraacetic acid (EDTA) was used for
masking interfering ions