پديد آورندگان :
Maleki N. نويسنده , SAFAVI A. نويسنده , Hasanpour F. نويسنده Department of Entomology, Faculty of Agriculture, Islamic Azad University of Arak , Shahbaazi H. R. نويسنده
چكيده لاتين :
Sensitive determination of 2,4-dichlorophenoxy acetic acid (2,4-0) was studied at a static mercury
drop electrode in aqueous media and in the presence of Britton-Robinson buffer as the electrolyte.
In this study, the electroinactive 2,4-0 was first converted to electroactive nitrated 2,4-0.
Subsequent determination was carried out by differential pulse polarography. Under optimum
conditions (pH = 11; v = 90 mV S^-1; modulation amplitude = 120 mV). the relative standard
deviation was 4.5% and limit of detection in terms of nitrated 2,4-D was 5 (MU)g L^-1 (1.88 X 10-8 M)
with linear concentration range from 8 to 300 /.1g L^-1(3 X 10^-8 -1.12 x 1O^-6M). The proposed
method offered high sensitivity as well as good selectivity. A mechanistic study, showed fourelectron
reduction of nitrated 2,4-0 to its hydroxylamine derivative. The proposed method was,
also, applied to the determination of 2,4-0 in environmental samples such as soil samples.