The existing C-30 cyclotron at NRCAM is currently capable of emitting a proton beam with only
Emax=30 MeV. To improve this Emax to higher energies of 250 MeV for proton therapy, this beam is injected
into the conceptually designed Medical Proton Synchrotron accelerator. Prior to injection, the beam
emittances in both x- and y-directions were measured to be varebsilon x=1.76„e10-5 m.rad and varebsilon y = 5.64„e10-5 m.rad. The
obtained extracted Twiss parameters at the point of injection were; alpha x =-1.28, beta x =12.7, gamma x =0.21,alpha y =-0.55,beta y
=3.28, and gamma y=0.10. The AGILE computer program code was utilized for the lattice design of the Medical
Proton Synchrotron. This design is composed of two achromatic arcs. Each arc bends 180 degree X and has a closed
dispersion bump. The designed lattice is composed of eight cells each consisting of two Quadrupoles, one
Sextupole and two Dipole magnets. The extraction components of sextupole resonance (SR), electrostatic
septum (ES) and magnetic septum (MS) were also placed in the designed lattice. The beta functions, as the
main parameters of the lattice, have been chosen to meet requirements of the slow extraction. The working
point (Qx, Qy) for adjusting the slow extraction were 1.33 and 2.62, respectively. The Hardt condition was met
by minimizing the angular dispersion of the separatists corresponding to a different energy spread. The slow
extraction of the accelerated particles was also analyzed. The calculation of the slow extraction is performed
using the AGILE program. The obtained lattice Twiss parameters of the focusing structure are also presented.