Author/Authors :
İyisoy, Atila Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Kurşaklıoğlu, Hürkan Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Çelik, Turgay Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Barçın, Cem Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Çelik, Murat Gulhane Military Medical Faculty - Department of Cardiology, Turkey
Title Of Article :
Retrieval of the Amplatzer duct occluder II device embolizing to the pulmonary artery by Goose Neck snare
شماره ركورد :
19865
Abstract :
Percutaneous transcatheter closure of patent ductus arteriosus is a wellestablished procedure. Although it has been known as a safe treatment strategy, some complications have been reported. Embolization of the device, especially to the pulmonary artery, is one of these complications. In this report, we present a patient in whom an Amplatzer® duct occluder (ADO) II device embolized to the pulmonary artery and it was retrieved by using “Goose Neck” snare. We would like to notice the great importance of accurate preprocedural measurement of ductus arteriosus diameter during percutaneous transcatheter closure which seems to be a simple procedure.
From Page :
276
NaturalLanguageKeyword :
Amplatzer duct occluder , “Goose Neck” snare , patent ductus arteriosus , percutaneous transcatheter closure
JournalTitle :
Gulhane Medical Journal
To Page :
279
Link To Document :
بازگشت