• Author/Authors

    seghiri, fatima university abou bakr belkaid - departement of physics, Tlemcen, Algeria , aliane, kamila preparatory school of oran - department of electrical engineering and electronics, Oran, Algeria , benabdallah, nadia preparatory school of tlemcen - departement of physics, Tlemcen, Algeria , lachachi, djamal university abou bakr belkaid - department of electrical engineering and electronics, Tlemcen, Algeria , benahmed, nasreddine university abou bakr belkaid - department of electrical engineering and electronics, Tlemcen, Algeria

  • Title Of Article

    Optimization and design of MRI birdcage probe with high-Q using eight-coupled microstrip resonator

  • شماره ركورد
    35427
  • Abstract
    This work describes the study of coil for Magnetic Resonance Imaging (MRI) application. Concretely, the principal objective is the design of the highest Q birdcage coil using the eight- coupled microstrip resonator and operating at 200 MHz. The numerical tool has been employed to analyze the inductance and capacitance matrices and the quality factor in terms of the thickness of the microstrip line.
  • From Page
    1
  • NaturalLanguageKeyword
    Birdcage , MRI , Analysis and design , MoM method , S , parameters , high , Q , Frequency response
  • JournalTitle
    Mediterranean Telecommunications Journal
  • To Page
    5
  • JournalTitle
    Mediterranean Telecommunications Journal