DocumentCode
10022
Title
Open platform, 32-channel, portable, data-logger with 32 pga control lines for wearable medical device development
Author
Gargiulo, Gaetano D. ; Bifulco, P. ; Cesarelli, M. ; McEwan, Alistair ; Wabnitz, A.
Author_Institution
MARCS Inst., Univ. of Western Sydney, Sydney, NSW, Australia
Volume
50
Issue
16
fYear
2014
fDate
July 31 2014
Firstpage
1127
Lastpage
1129
Abstract
A compact isolated low-power 32-channel 16-bit data-logging system around an NXP ARM processor (LPC1768) and four of the linear technology octal analogue-to-digital converters (ADCs) LTC1857/58/59 is designed. The system requires only 250 mA when powered at 5 V to run at full power (including a capacitive 2.8 inch touch-screen display and 32 Gb SHDC SD memory card). The sample rate is configurable up to 1 k SPS per channel as well as voltage dynamic input up to ±10 V; additionally, 32 chip select lines (SPI protocol) individually addressable and controllable while sampling to configure user-designed programmable gain amplifiers (PGAs) are available. Collaboration is being sought to improve the software capabilities, particularly to enrich the very basic user interface and to add wireless connectivity. The code is available (under the GPL licence) at our repository, the gerber file to reproduce the PCB is available (on the As-Is basis) on request. The galvanic isolation between the power supply data connection and ADC channels makes the data-logger also compatible with the main powered PCs, hence it is suitable for the implementation of medical devices at least for the prototyping and initial testing stages.
Keywords
amplifiers; analogue-digital conversion; biomedical electronics; biomedical equipment; data loggers; low-power electronics; microprocessor chips; power supply circuits; printed circuits; programmable circuits; ADC; ADC channel; GPL licence; LPC1768; LTC1857-58-59; NXP ARM processor; PCB; PGA control line; SHDC SD memory card; SPI protocol; current 250 mA; galvanic isolation; linear technology octal analog-to-digital converter; low-power portable data-logging system; power supply data connection; storage capacity 32 Gbit; touch-screen display; user-designed programmable gain amplifier; voltage 5 V; wearable medical device development; word length 16 bit;
fLanguage
English
Journal_Title
Electronics Letters
Publisher
iet
ISSN
0013-5194
Type
jour
DOI
10.1049/el.2014.1791
Filename
6870594
Link To Document