Title :
Pulse parameter dependence on transition occurrence instant and waveform epoch
Author :
Larson, Donald R. ; Paulter, Nicholas G. ; Bergman, David I.
Author_Institution :
Quantum Electr. Metrol. Div., Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Abstract :
The pulse parameters of amplitude and transition duration are dependent on the waveform epoch and transition occurrence instant of the pulse transition in the waveform epoch. The primary explanations for the observed variations are the pulse aberrations and settling behavior of both the pulse generator and the measurement instrument (sampling oscilloscope). Measurement results are included for three pulse generators and two sampling oscilloscopes.
Keywords :
oscilloscopes; pulse generators; pulse measurement; waveform analysis; measurement instrument; pulse aberrations; pulse generator; pulse parameter; pulse transition; sampling oscilloscope; settling error; transition duration; waveform epoch; Amplitude estimation; Attenuation; Bandwidth; Calibration; Instruments; NIST; Oscilloscopes; Pulse generation; Pulse measurements; Sampling methods; Picosecond; pulse parameter; sampling oscilloscope; settling error; transition duration;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :