• DocumentCode
    1008981
  • Title

    Performance of 2 mm radius straw tube drift cells

  • Author

    Avery, R.E. ; Bantly, J. ; Blessing, S. ; Buchholz, D. ; Gobbi, B. ; Liu, Y. ; Rajagopalan, S. ; Tilden, R. ; Martin, M.

  • Author_Institution
    Northwestern Univ., Evanston, IL, USA
  • Volume
    40
  • Issue
    4
  • fYear
    1993
  • fDate
    8/1/1993 12:00:00 AM
  • Firstpage
    578
  • Lastpage
    582
  • Abstract
    The performance of a 128-channel test module made with straw tubes of 2-mm radius was studied in a test beam and with cosmic rays. Different gases were used, and for each one the time-to-distance relation and the hit efficiency were measured. Comparisons are made between results obtained using two different electronics readouts. The best solution, of 135 μm, was obtained using 50% ethane, 50% argon and reading out the information with the TDCs (time-to-digital converters), at an operating voltage of 1750 V
  • Keywords
    position sensitive particle detectors; radiation detection and measurement; 135 micron; 1750 V; 2 mm; Ar ethane mixture; TDCs; cosmic rays; hit efficiency; readout; straw tube drift cells; time-to-digital converters; time-to-distance relation; Aluminum; Cosmic rays; Counting circuits; Detectors; Gases; Laboratories; Structural beams; Testing; Voltage; Wires;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.256620
  • Filename
    256620