DocumentCode
1009827
Title
Imaging of radionuclide emissions with a low-noise charge-coupled device
Author
Karellas, Andrew ; Liu, Hong ; Reinhardt, Christopher ; Harris, Lisa J. ; Brill, A.B.
Author_Institution
Massachusetts Univ., Med. Center, Worcester, MA, USA
Volume
40
Issue
4
fYear
1993
fDate
8/1/1993 12:00:00 AM
Firstpage
979
Lastpage
982
Abstract
The adaptation of a cooled charged-coupled device (CCD) for some autoradiographic applications which may alleviate some of the problems encountered with film is investigated. The proposed approach uses a scientific-grade CCD which is optically coupled via a lens to a scintillator without intermediate amplification. On the basis of experience with lens coupling, the performance of the CCD is investigated with the option of a fiberoptic coupling between the scintillator and CCD. The attainable performance by operating the CCD in the charge integration mode is explored experimentally, and computations are presented in order to predict the performance of a modified imaging system operating in the counting pulse-height mode for energy discrimination
Keywords
CCD image sensors; biomedical equipment; discriminators; pulse height analysers; radioisotope scanning and imaging; scintillation counters; CCD; autoradiographic applications; charge integration mode; cooled charged-coupled device; counting pulse-height mode; energy discrimination; fiberoptic coupling; film; lens coupling; low-noise charge-coupled device; modified imaging system; optically coupled; radionuclide emissions; scientific-grade; scintillator; Charge coupled devices; Charge-coupled image sensors; Dynamic range; Electrons; Lenses; Optical coupling; Optical films; Optical imaging; Optical sensors; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.256696
Filename
256696
Link To Document