• DocumentCode
    1010714
  • Title

    The energizing of a NMR superconducting coil by a superconducting rectifier

  • Author

    Sikkenga, J. ; Knoben, J. ; Spoorenberg, C. J G ; Kraaij, G.J. ; Klundert, L. J M V D ; Kate, H. H J Ten

  • Author_Institution
    HOLEC Innovation and Technology, Hengelo, The Netherlands
  • Volume
    21
  • Issue
    2
  • fYear
    1985
  • fDate
    3/1/1985 12:00:00 AM
  • Firstpage
    440
  • Lastpage
    443
  • Abstract
    NMR magnets require a good homogeneity within a certain volume and an excellent field stability. The homogeneity can be met using a superconducting shim coil system. The field stability requires a constant current, although in many cases the current decay time constant is too low, due to imperfections in the superconducting wire and joints. This can be overcome using a rectifier. The rectifier can also be used to load the coil. The combination and interaction of the superconducting NMR coil (2.0 Tesla and 0.35 m cold bore) and the rectifier (20 W / 1 kA) is tested. The safety of the system is discussed. The shim coil system can compensate the strayfield of the rectifier. The field decay compensation will be discussed.
  • Keywords
    Nuclear magnetic resonance imaging; Rectifiers; Superconducting coils; Boring; Nuclear magnetic resonance; Rectifiers; Safety; Stability; Superconducting coils; Superconducting filaments and wires; Superconducting magnets; Switches; Technological innovation;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.1985.1063857
  • Filename
    1063857