Title :
Spectroscopic Measurements of Oxygen Saturation in the Retina
Author :
Ramella-Roman, Jessica C. ; Mathews, Scott A.
Author_Institution :
Catholic Univ. of America, Washington
Abstract :
We introduce a new type of retinal oximeter able to capture multiple spectroscopic sensitive images of the retina in a single snapshot. The core of our apparatus is a multiaperture camera that can be easily interfaced with a fundus camera system. A lenslet array divides the image collected by a fundus camera into identical subimages. A bandpass filter array in front of the lens system is used to collect the subimages only at the wavelengths of interest. Two different instruments were designed based on this simple principle. The first system can collect six different images, while the second system can collect up to 18 different images in a single snapshot.
Keywords :
band-pass filters; biomedical optical imaging; cameras; eye; image registration; image sensors; oximetry; spectrochemical analysis; bandpass filter array; fundus camera system; lenslet array; multiaperture camera; multiple spectroscopic sensitive image capturing; oxygen saturation assessment; retinal oximeter; single snapshot image; spectroscopic measurements; Cameras; Diabetes; Optical films; Optical filters; Optical saturation; Optical scattering; Optical sensors; Oxygen; Retina; Spectroscopy; Image registration; optical spectroscopy;
Journal_Title :
Selected Topics in Quantum Electronics, IEEE Journal of
DOI :
10.1109/JSTQE.2007.911312