Title
Calibration of High-Voltage Current-Comparator-Based Capacitance and Dissipation Factor Bridges Using Nonsynchronous Sampling
Author
Kyriazis, Gregory A.
Author_Institution
Normalizacao e Qualidade Ind., Inst. Nac. de Metrologia, Caxias
Volume
58
Issue
4
fYear
2009
fDate
4/1/2009 12:00:00 AM
Firstpage
1029
Lastpage
1034
Abstract
An alternative approach for calibrating automated high-voltage current-comparator-based capacitance and dissipation factor bridges is presented. For the calibration, two sinusoidal currents are synthesized with two synchronized signal generators and two resistors and injected into the bridge inputs. The capacitance ratio and dissipation factor simulated by these currents are calculated from an ac voltage ratio estimated from the digitized data, together with the impedance ratio of the resistors. The approach is a refinement of another approach previously published that leads to larger uncertainties but is slightly easier to implement since the digitizers need not be synchronized to the signal generators. It is possible to simulate capacitance ratios ranging from 1 : 1 to 10 : 1 with relative standard uncertainties of less than 2.0 middot10-5 and dissipation factors ranging from 0 to 0.1 with standard uncertainties of less than 1 middot10-5.
Keywords
bridge circuits; calibration; capacitance; current comparators; signal generators; automated high-voltage current-comparator; capacitance ratio; dissipation factor bridges; impedance ratio; nonsynchronous sampling; sinusoidal currents; synchronized signal generators; AC bridge; calibration; capacitance; digital voltmeters; dissipation factor; sampling methods;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2008.2008082
Filename
4694095
Link To Document