• DocumentCode
    1024535
  • Title

    Needle Enzyme Electrode for Lactate Measurement In Vivo

  • Author

    Rong, Zimei ; Leitao, Eugenia ; Popplewell, Jonathan ; Alp, Burçak ; Vadgama, Pankaj

  • Author_Institution
    Queen Mary Univ. of London, London
  • Volume
    8
  • Issue
    1
  • fYear
    2008
  • Firstpage
    113
  • Lastpage
    120
  • Abstract
    Electrochemical lactate needle enzyme electrodes were fabricated based on lactate oxidase with a conventional hydrogen peroxide detection regimen with a linear range up to 7 mM, response time ~ 3 min, and sensitivity ~ 1 nA/mM. A negatively charged inner (sulphonated polyether ether sulphone-polyether sulphone) membrane was applied for ensuring selectivity by limiting oxidazible anion diffusion to the Pt working electrode; polyurethane outer membrane layers were dip coated over the enzyme layer to limit substrate diffusion to the enzyme layer to achieve: (1) stir independence and (2) a low oxygen requirement. Lactate was monitored subcutaneously in rats during controlled haemorrhage and hypovolaemic shock. Tissue lactate showed agreement with blood lactate before haemorrhage and for limited haemorrhage (up to 2 ml blood withdrawal from 16 ml total blood volume) but with blood loss above 3 ml the tissue lactate rise was less pronounced than in blood. Loss of intercompartmental equilibrium due to diffusion limitation is suggested as a factor in causing this difference. An experimental in vitro model was developed which employed the needle lactate electrode within a cylindrical collagen gel to monitor inward diffusion of lactate as a basis for determining lactate diffusion coefficient. The high precision measurement gave a diffusion coefficient consistent with report values 3.54 times 10-6 cm2/s. The simplified experimental approach could allow lactate transport studies across tissue analogues.
  • Keywords
    biochemistry; biomedical measurement; biosensors; electrochemical electrodes; electrochemical sensors; enzymes; molecular biophysics; biosensors; electrochemical electrodes; enzyme layer; haemorrhage shock; hydrogen peroxide detection; hypovolaemic shock; intercompartmental equilibrium loss; lactate diffusion coefficient; lactate measurement; lactate oxidase; needle enzyme electrode; negatively charged inner membrane; oxidazible anion diffusion; sulphonated polyether ether sulphone-polyether sulphone; Biochemistry; Biomembranes; Blood; Delay; Electrodes; Hemorrhaging; In vivo; Monitoring; Needles; Rats; Animal; biosensors; blood; computer-aided analysis; lactate; measurement; subcutaneous;
  • fLanguage
    English
  • Journal_Title
    Sensors Journal, IEEE
  • Publisher
    ieee
  • ISSN
    1530-437X
  • Type

    jour

  • DOI
    10.1109/JSEN.2007.912919
  • Filename
    4418372