DocumentCode
1024631
Title
Considerations in the design of MRI magnets with reduced stray fields
Author
Hawksworth, D.G. ; McDougall, I.L. ; Bird, J.M. ; Black, D.
Author_Institution
Oxford Magnet Technology Ltd., Oxon, England
Volume
23
Issue
2
fYear
1987
fDate
3/1/1987 12:00:00 AM
Firstpage
1309
Lastpage
1314
Abstract
One of the major considerations in siting choice for an MRI system within a hospital is its interactions with its environment. This interaction places restrictions on the proximity of equipment sensitive to magnetic field and limits areas of general public access. In addition special account must be taken of the possible impact of environmental iron on magnet homogeneity. To date, the approach adopted by the MRI system scanner manufacturers to these problems has been to employ either YOKE or MIRROR iron (PASSIVE) shielding frequently requiring significant structural modifications with associated costs. An alternative approach to the shielding problem has been investigated at OXFORD using a geometry utilising superconducting counter running coils alone and prototypes with Central fields of 0.5T, 1.0T and 1.5T have been tested.
Keywords
Biomedical imaging, MRI; Magnetic shielding; Superconducting magnets; Costs; Geometry; Hospitals; Iron; Magnetic resonance imaging; Magnetic shielding; Magnets; Manufacturing; Mirrors; Superconducting coils;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.1987.1065051
Filename
1065051
Link To Document