Title : 
A heat fluxmeter based on the Nernst effect
         
        
        
            Author_Institution : 
Electr. Ind., Conservatoire Nat. des Arts et Metiers, Paris, France
         
        
        
        
        
            fDate : 
3/1/1990 12:00:00 AM
         
        
        
        
            Abstract : 
The use of a thermomagnetoelectric effect called the Nernst effect to reduce the intrinsic time constant of heat fluxmeters is proposed. The Nernst voltage was measured on a polycrystalline bismuth sample in terms of the heat flux, the magnetic induction field, and the temperature. A new heat fluxmeter, based on the Nernst effect and fitted to a 15-cm3 cell, was built and tested. At room temperature, its sensitivity is 0.6 μV/mW and the main intrinsic time constant is 25 s
         
        
            Keywords : 
bismuth; thermal variables measurement; thermomagnetic effects; Nernst effect; heat fluxmeter; intrinsic time constant; magnetic induction field; polycrystlaline Bi; room temperature; sensitivity; thermomagnetoelectric effect; Bismuth; Conductors; Copper; Detectors; Electrodes; Heat sinks; Temperature sensors; Thermal variables measurement; Thermoelectricity; Voltage;
         
        
        
            Journal_Title : 
Magnetics, IEEE Transactions on