DocumentCode :
1034740
Title :
Dead zone characteristics of SQS and GM tubes
Author :
Koori, N. ; Nohtomi, A. ; Hashimoto, M. ; Yoshioka, K. ; Kumabe, I. ; Matsubara, S. ; Yoshizumi, M. ; Oshima, T.
Author_Institution :
Dept. of Nucl. Eng., Kyushu Univ., Fukuoka, Japan
Volume :
37
Issue :
2
fYear :
1990
fDate :
4/1/1990 12:00:00 AM
Firstpage :
107
Lastpage :
111
Abstract :
The characteristics of the dead zone of a gas counter were studied at gas pressures of 760-150 torr of Ar(75)+isoC4H10(25) mixture and compared with those of a commercially available GM tube. The counter operated in the SQS (self-quenching streamer) mode at pressures higher than 350 torr, and in the GM mode at the lower pressures; the operation mode changed around 350 torr, where the mean free path of VUV (vacuum ultraviolet) photons is about 50 μm. The dead zone of the counter used was evaluated to be 290 μs-cm for the SQS mode (at 760 torr) and 2450 μs-cm for the GM mode (at 150 torr). The counting rate capability of the SQS tube is more than one order of magnitude larger than that of the GM tube
Keywords :
Geiger counters; gamma-ray detection and measurement; streamer chambers; commercially available GM tube; dead zone; gas counter; self-quenching streamer; Atmospheric measurements; Current measurement; Loss measurement; Pulse measurements; Pulse shaping methods; Radiation detectors; Shape measurement; Size measurement; Time measurement; X-rays;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.106600
Filename :
106600
Link To Document :
بازگشت