DocumentCode
1036663
Title
Experimental verification of theoretical predictions concerning the optimum defibrillation waveform
Author
Malkin, R.A. ; Jackson, S.R. ; Nguyen, John ; Zhao Yang ; Dongxu Guan
Author_Institution
Dept. of Biomed. Eng., Duke Univ., Durham, NC
Volume
53
Issue
8
fYear
2006
Firstpage
1492
Lastpage
1498
Abstract
The efficacy of electrical therapy at terminating ventricular fibrillation is highly dependent on the waveform used. We present experimental results which test one theory for defibrillation waveform dependence. Forty-four defibrillation waveforms (22 monophasic, 22 biphasic) were designed according to the theoretical construct of Fishler (2000). The waveforms were then tested on 67 male guinea pigs (46 for monophasic, 21 for biphasic waveforms) using a custom designed defibrillator and 12-mm subcutaneous disc electrodes. There was considerable agreement between the theoretical and experimental results. For example, as predicted, the ascending exponential waveform of 1 ms proved to be the most effective (86.4%) monophasic waveform, where efficacy is the number of successful shocks divided by the total number delivered. In addition, the efficacy decrease with duration increase was accurately predicted by the model for monophasic waveforms. For biphasic waveforms, as predicted by the model, when the first phase was optimized, an increase in second phase duration caused an increase in defibrillation efficacy (10 of 11 tested duration pairs). We conclude that the theoretical framework adequately explains the mechanism by which the defibrillation waveform affects efficacy for monophasic waveforms and, in at least one aspect, biphasic waveforms
Keywords
bioelectric phenomena; biomedical electrodes; defibrillators; electrocardiography; patient treatment; 1 ms; 12 mm; biphasic waveform; custom designed defibrillator; electrical therapy; guinea pigs; monophasic waveform; optimum defibrillation waveform; subcutaneous disc electrodes; ventricular fibrillation; Biomedical engineering; Defibrillation; Electric shock; Electrodes; Fibrillation; Medical treatment; Myocardium; Predictive models; Testing; Cardiac electrophysiology; defibrillation; Animals; Computer Simulation; Electric Countershock; Guinea Pigs; Heart Conduction System; Heart Ventricles; Male; Models, Cardiovascular; Therapy, Computer-Assisted; Treatment Outcome; Ventricular Fibrillation;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2006.876643
Filename
1658143
Link To Document