• DocumentCode
    1037306
  • Title

    A system for precise determination of effective atomic number by beta backscatter

  • Author

    Soller, M.S. ; Cirignano, L. ; Lieberman, P. ; Squillante, M.R.

  • Author_Institution
    Radiat. Monitoring Devices, Inc., Watertown, MA, USA
  • Volume
    37
  • Issue
    2
  • fYear
    1990
  • fDate
    4/1/1990 12:00:00 AM
  • Firstpage
    230
  • Lastpage
    233
  • Abstract
    A compact instrument capable of determining the effective atomic number (Zeff) of a target has been developed and characterized. It is a nondestructive sensor which utilizes a cadmium telluride solid-state detector and a 90Sr/90Y source in a backscatter geometry. The spectrum of the backscattered beta particles depends on the target atomic number. Measurements show that, by using the mean energy of the backscatter spectrum, the average effective Atomic number can be determined to better than ±1Z . The system includes data acquisition and analysis software which runs on a personal computer. The system was designed to be flexible enough to be used for a variety of applications by changing the data analysis algorithms and the source type and energy
  • Keywords
    cadmium compounds; data acquisition; nuclear electronics; semiconductor counters; 90Sr/90Y source; CdTe; beta backscatter; data acquisition; data analysis algorithms; effective atomic number; nondestructive sensor; personal computer; Atomic measurements; Backscatter; Cadmium compounds; Data analysis; Detectors; Geometry; Instruments; Sensor phenomena and characterization; Solid state circuits; Strontium;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.106624
  • Filename