• DocumentCode
    1038598
  • Title

    Contact-Based Stiffness Sensing of Human Eye

  • Author

    Kurita, Yuichi ; Kempf, Roland ; Iida, Yoshichika ; Okude, Jumpei ; Kaneko, Makoto ; Mishima, Hiromu K. ; Tsukamoto, Hidetoshi ; Sugimoto, Eiichiro ; Katakura, Seiki ; Kobayashi, Ken ; Kiuchi, Yoshiaki

  • Author_Institution
    Nara Inst. of Sci. & Technol. (NAIST), Nara
  • Volume
    55
  • Issue
    2
  • fYear
    2008
  • Firstpage
    739
  • Lastpage
    745
  • Abstract
    Goldmann applanation tonometry is commonly used for measuring intraocular pressure (IOP) to diagnose glaucoma. However, the measured IOP by conventional applanation tonometry is valid only under the assumption that all subjects have the same structural eye stiffness. This paper challenges in vivo measurement of eye stiffness with a noninvasive approach and investigates individual differences of eye stiffness. Eye stiffness is defined by the applied force and displacement of the cornea. The displacement is detected based on captured images by a high resolution camera. The experimental results show that the measured stiffness nicely matches the analytical result that is derived from a simple spherical deformation model with an internal pressure. However, some subjects have different eye stiffness even with the same IOP. IOP with abnormal stiffness may be over/underestimated by conventional applanation tonometry. The proposed eye stiffness measurement can help detect the misestimated eye and it contributes to the early detection of glaucoma.
  • Keywords
    biomechanics; biomedical measurement; biomedical optical imaging; deformation; diseases; displacement measurement; elasticity; eye; Goldmann applanation tonometry; applied cornea force; contact-based stiffness sensing; cornea displacement; glaucoma diagnosis; high resolution camera; human eye; image capture; in vivo eye stiffness measurement; intraocular pressure measurement; spherical deformation model; structural eye stiffness; Biomedical measurements; Cornea; Eyes; Hospitals; Humans; Image resolution; In vivo; Medical treatment; Pressure measurement; Railway engineering; Contact method; glaucoma; human eye; stiffness sensing; Computer Simulation; Cornea; Diagnosis, Computer-Assisted; Elasticity; Humans; Intraocular Pressure; Models, Biological; Reproducibility of Results; Sensitivity and Specificity; Stress, Mechanical; Tonometry, Ocular;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2007.912650
  • Filename
    4432742