• DocumentCode
    1038656
  • Title

    Distribution of Electrical Stimulation Current in a Planar Multilayer Anisotropic Tissue

  • Author

    Mesin, Luca ; Merletti, Roberto

  • Author_Institution
    Politecnico di Torino, Torino
  • Volume
    55
  • Issue
    2
  • fYear
    2008
  • Firstpage
    660
  • Lastpage
    670
  • Abstract
    This study analytically addresses the problem of neuromuscular electrical stimulation for a planar, multilayer, anisotropic model of a physiological tissue (referred to as volume conductor). Both conductivity and permittivity of the volume conductor are considered, including dispersive properties. The analytical solution is obtained in the 2- D Fourier transform domain, transforming in the planes parallel to the volume conductor surface. The model is efficient in terms of computational cost, as the solution is analytical (only numerical Fourier inversion is needed). It provides the current distribution in a physiological tissue induced by an electrical current delivered at the skin surface. Three representative examples of application of the model are considered. (1) The simulation of stimulation artefact during transcutaneous electrical stimulation and EMG detection. Only the effect of the volume conductor is considered, neglecting the other sources of artefact (such as the capacitive coupling between the stimulating and recording electrodes). (2) The simulation of the electrical current distribution within the muscle and the low-pass filter effect of the volume conductor on sinusoidal stimulation currents with different stimulation frequencies. (3) The estimation of the amplitude modulated current distribution within the muscle for interferential stimulation. The model is devoted to the simulation of neuromuscular stimulation, but the same method could be applied in other fields in which the estimation of the electrical current distribution in a medium induced by the injection of a current from the boundary of the medium is of interest.
  • Keywords
    Fourier transforms; bioelectric phenomena; biological tissues; current distribution; electrical conductivity; electromyography; neuromuscular stimulation; permittivity; EMG detection; Fourier transform domain; amplitude modulated current distribution; anisotropic physiological tissue model; dispersive properties; electrical conductivity; electrical stimulation current distribution; interferential stimulation; interferential therapy; low-pass filter effect; neuromuscular electrical stimulation; numerical Fourier inversion; permittivity; planar multilayer anisotropic tissue; stimulus artefact; transcutaneous electrical stimulation; volume conductor; Anisotropic magnetoresistance; Conductivity; Conductors; Current distribution; Electrical stimulation; Frequency estimation; Muscles; Neuromuscular; Nonhomogeneous media; Permittivity; Electrical stimulation; interferential therapy; stimulus artefact; Anisotropy; Computer Simulation; Electric Stimulation; Humans; Models, Biological; Motor Neurons; Muscle Contraction; Muscle, Skeletal; Skin Physiology;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2007.902248
  • Filename
    4432747