    Quantitative Rotating Multisegment Slant-Hole SPECT Mammography With Attenuation and Collimator-Detector Response Compensation

  • Author

    Xu, Jingyan ; Liu, Chi ; Wang, Yuchuan ; Frey, Eric C. ; Tsui, Benjamin M W

  • Author_Institution
    Johns Hopkins Univ., Baltimore
  • Volume
    26
  • Issue
    7
  • fYear
    2007
  • fDate
    7/1/2007 12:00:00 AM
  • Firstpage
    906
  • Lastpage
    916
  • Abstract
    Rotating multisegment slant-hole (RMSSH) single photon emission computed tomography (SPECT) is suitable for detecting small and low-contrast breast lesions since it has much higher detection efficiency than conventional SPECT with a parallel-hole collimator and can image the breast at a closer distance. Our RMSSH SPECT reconstruction extends a previous rotation-shear transformation-based method to include nonuniform attenuation and collimator-detector response (CDR) compensation. To evaluate our reconstruction method, we performed two phantom simulation studies with 1) an isolated breast and 2) a breast phantom attached to the body torso. The reconstructed RMSSH SPECT images with attenuation and CDR compensation showed improved quantitative accuracy and less image artifacts than without. To evaluate the clinical efficacy of RMSSH SPECT mammography, we used a simulation study to compare with planar scintimammography in terms of the signal-to-noise ratio (SNR) value of a breast lesion. The RMSSH SPECT reconstruction images showed higher SNR value than the planar scintimammography images and even more so as we applied compensation for attenuation and collimator detector response. We conclude that attenuation and CDR compensation provide RMSSH SPECT mammography images with improved quality and quantitative accuracy.
  • Keywords
    biomedical imaging; cancer; image reconstruction; mammography; single photon emission computed tomography; breast phantom; collimator-detector response compensation; isolated breast; low-contrast breast lesions; parallel-hole collimator; planar scintimammography; rotating multisegment slant-hole SPECT mammography; rotation-shear transformation-based method; signal-to-noise ratio; single photon emission computed tomography; Attenuation; Breast; Image reconstruction; Imaging phantoms; Lesions; Mammography; Optical collimators; Performance evaluation; Reconstruction algorithms; Single photon emission computed tomography; Algorithms; Artifacts; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Mammography; Phantoms, Imaging; Reproducibility of Results; Rotation; Sensitivity and Specificity; Tomography, Emission-Computed, Single-Photon;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2007.895480
  • Filename
    4265758