• DocumentCode
    1050555
  • Title

    Reconstructing current distributions from biomagnetic measurements under large external noise disturbances

  • Author

    Sekihara, Kensuke ; Scholz, Bernhard

  • Author_Institution
    Central Res. Lab., Hitachi Ltd, Tokyo
  • Volume
    13
  • Issue
    1
  • fYear
    1994
  • fDate
    3/1/1994 12:00:00 AM
  • Firstpage
    144
  • Lastpage
    151
  • Abstract
    External noise fields cause spatially coherent noise in the biomagnetic data measured by a multichannel magnetometer. The authors propose a method of incorporating this spatial coherence into current-density reconstruction. This method can reconstruct current distributions from biomagnetic measurements affected by external noise fields. Computer simulations demonstrate its effectiveness
  • Keywords
    bioelectric phenomena; biomagnetism; image reconstruction; medical image processing; noise; biomagnetic imaging; biomagnetic measurements; computer simulations; current distributions reconstruction; current-density reconstruction; large external noise disturbances; medical imaging; multichannel magnetometer; spatial coherence; Biomagnetics; Coils; Computer simulation; Current measurement; Detectors; Image reconstruction; Magnetic field measurement; Magnetic noise; Noise measurement; Working environment noise;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.276153
  • Filename
    276153