• DocumentCode
    1050605
  • Title

    High peak SAR exposure unit with tight exposure and environmental control for in vitro experiments at 1800 MHz

  • Author

    Schuderer, Jürgen ; Samaras, Theodoros ; Oesch, Walter ; Spät, Denis ; Kuster, Niels

  • Author_Institution
    Integrated Syst. Lab., Swiss Fed. Inst. of Technol., Zurich, Switzerland
  • Volume
    52
  • Issue
    8
  • fYear
    2004
  • Firstpage
    2057
  • Lastpage
    2066
  • Abstract
    The objective of this study was to develop, optimize, and characterize a flexible and reliable unit for standardized and well-controlled electromagnetic-field exposure of cells in vitro. The technical requirements were high peak and time-averaged exposure of the cells with a minimal temperature rise for the cell cultures, flexible modulation schemes, high uniformity, and low variability of exposure, as well as support of blinded protocols. The developed setup is based on two R18 waveguides resonant at 1800 MHz and operated with a computer-controlled signal and monitoring unit. The cells can be exposed in 35-mm Petri dishes either cultivated as monolayers or in suspension. For cell monolayers, the system provides an efficiency for the specific absorption rate (SAR) of >50 (W/kg)/W input power, a nonuniformity of the SAR distribution of <30%, SAR variability of <6%, and a temperature rise of <0.03°C/(W/kg) average SAR. For cell suspensions and provided that the cells are not in the meniscus area, a SAR efficiency of >10 (W/kg)/W, nonuniformity of <40%, SAR variability of <17%, and a temperature rise of <0.13°C/(W/kg) is achieved. The numerical dosimetry for the field and temperature distributions within the Petri dishes was verified using E-field and temperature probes. The temperature analysis has shown that the possibility of localized "hot spots" can be excluded.
  • Keywords
    biological effects of microwaves; biothermics; cellular effects of radiation; computerised control; computerised monitoring; dosimetry; electromagnetic wave absorption; finite difference time-domain analysis; mobile communication; monolayers; suspensions; waveguide couplers; 1800 MHz; Petri dish; R18 waveguides; SAR exposure; blinded protocols; cell cultures; cell monolayers; computer-controlled signal; electromagnetic-field exposure; environmental control; flexible modulation; in vitro experiments; localized hot spots; monitoring unit; monolayers; numerical dosimetry; specific absorption rate; suspension; temperature probes; Cells (biology); Electromagnetic waveguides; GSM; In vitro; Laboratories; Plasma temperature; Protocols; Radio frequency; Specific absorption rate; Temperature distribution; Dosimetry; RF; SAR; exposure setup; specific absorption rate;
  • fLanguage
    English
  • Journal_Title
    Microwave Theory and Techniques, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9480
  • Type

    jour

  • DOI
    10.1109/TMTT.2004.832009
  • Filename
    1318804