DocumentCode :
105320
Title :
Integration of TES Microcalorimeters With Microwave SQUID Multiplexed Readout
Author :
Bennett, Douglas A. ; Mates, John A. B. ; Gard, Johnathon D. ; Hoover, Andrew S. ; Rabin, Michael W. ; Reintsema, Carl D. ; Schmidt, Daniel R. ; Vale, Leila R. ; Ullom, Joel N.
Author_Institution :
Quantum Electron. & Photonics Div., Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Volume :
25
Issue :
3
fYear :
2015
fDate :
Jun-15
Firstpage :
1
Lastpage :
5
Abstract :
The demonstration of a microwave superconducting quantum interference device (SQUID) multiplexed readout of transition-edge sensor (TES) microcalorimeters has the potential to dramatically expand the scale of arrays of TESs. In this paper, we discuss recent work to develop an instrument for high-resolution high-efficiency gamma-ray spectroscopy that integrates previously demonstrated high-resolution TES microcalorimeters with new lower noise microwave SQUID multiplexers. We will discuss the proposed instrument design and readout noise of the optimized microwave SQUID multiplexed readout. Finally, we will discuss the potential limits on the number of multiplexed TESs imposed by the analog-to-digital converter.
Keywords :
SQUIDs; calorimeters; gamma-ray spectroscopy; superconducting microwave devices; TES microcalorimeters; analog-to-digital converter; high-resolution high-efficiency gamma-ray spectroscopy; microwave SQUID multiplexed readout; microwave superconducting quantum interference device multiplexed readout; transition-edge sensor microcalorimeters; HEMTs; Instruments; Microwave photonics; Multiplexing; SQUIDs; Signal to noise ratio; SQUIDs; Superconducting microwave devices; Superconducting photodectectors; superconducting microwave devices; superconducting photodectectors;
fLanguage :
English
Journal_Title :
Applied Superconductivity, IEEE Transactions on
Publisher :
ieee
ISSN :
1051-8223
Type :
jour
DOI :
10.1109/TASC.2014.2381878
Filename :
6994798
Link To Document :
بازگشت