DocumentCode
1058891
Title
Optimization Method for TES Microcalorimeters With Absorbers
Author
Oshima, Tai ; Yoshitake, Hiroshi
Author_Institution
Nobeyama Radio Obs., Nat. Astron. Obs. of Japan (NRO/NAOJ), Nagano, Japan
Volume
19
Issue
3
fYear
2009
fDate
6/1/2009 12:00:00 AM
Firstpage
465
Lastpage
468
Abstract
Large absorbers and a high counting rate is required for high energy resolution transition-edge sensor (TES) microcalorimeters for future X-ray astronomical missions. However, these two requirements have a relation of a trade-off in terms of energy resolution. Thus, study of the effect of thermal diffusion in large absorbers on energy resolution is essential. Therefore, we performed 3 dimensional FEM simulations for TES microcalorimeter with Bi/Au absorber, where a role of Au layer is boost the thermalization of slow thermalizing Bi, using SPICE to calculate the effect of thermal diffusion in the absorber on energy resolution. The simulation results for square absorbers of size 500 mum and 750 mum with Au thickness 0.5 mum indicate that energy resolution of 3 eV and 5 eV can be achieved, respectively, with a fast response time of 100 mus, provided the residual resistivity ratio of Au is higher than ~10.
Keywords
bismuth; calorimeters; finite element analysis; gold; optimisation; superconducting particle detectors; thermal diffusion; 3 dimensional FEM simulations; Bi-Au; SPICE; TES microcalorimeters; absorbers; energy resolution; optimization method; residual resistivity ratio; size 0.5 mum; size 500 mum; size 750 mum; thermal diffusion; thermalization; transition-edge sensor; Microcalorimeter; superconducting transition edge sensors;
fLanguage
English
Journal_Title
Applied Superconductivity, IEEE Transactions on
Publisher
ieee
ISSN
1051-8223
Type
jour
DOI
10.1109/TASC.2009.2019568
Filename
5067040
Link To Document