Title :
Calibration of time domain network analyzers
Author :
Su, Wansheng ; Riad, Sedki M.
Author_Institution :
Bradley Dept. of Electr. Eng., Virginia Polytech. Inst. & State Univ., Blacksburg, VA, USA
fDate :
4/1/1993 12:00:00 AM
Abstract :
A calibration technique for a time domain reflectometry and transmission (TDR and TDT) measurement system as applied to network analysis is presented. The calibration corrects for the errors caused by the response of the measurement system. A complete physically-based model is established for the system. A set of calculable standards is developed to satisfy the time domain requirements for calibration. The calibration technique was applied to determining the model parameters of a commercial TDR and TDT system. The calibration enhanced the system bandwidth from 8 GHz to about 20 GHz. Experimental verification is given to demonstrate the validity and accuracy of the calibration technique
Keywords :
calibration; measurement standards; microwave reflectometry; network analysers; time-domain reflectometry; 8 to 20 GHz; TDR; TDT; calibration; measurement standards; network analysis; standards; time domain network analyzers; time domain reflectometry; Bandwidth; Calibration; Impedance measurement; Measurement standards; Oscilloscopes; Pulse generation; Sampling methods; Time domain analysis; Time measurement; Transmission line measurements;
Journal_Title :