• DocumentCode
    1059233
  • Title

    Establishing a Josephson-array voltage standard system at NML, ROC

  • Author

    Hwang, Gwo-Jen ; Chien, Shih-Sen ; Chen, Chun-Chun ; Ma, Li-Jehn ; Chan, Jiing-Ching

  • Author_Institution
    Nat. Meas. Lab., Centere for Meas. Stand., Hsinchu, Taiwan
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    227
  • Lastpage
    230
  • Abstract
    A 1-V Josephson-array voltage standard system has been operational at the National Measurement Laboratory (NML) of the Center for Measurement Standards (CMS) in the Republic of China (ROC). To help evaluate its accuracy, the authors selected two Zener references that had been calibrated by means of this system over a three-month period and hand-carried them to four other national laboratories around the world for comparison purposes. Over a four-week period in November and December, 1991, these two references were measured in turn at PTB (Germany), NPL (UK), NIST (USA), and ETL (Japan). The results of these international comparisons at the 1-V level, using the Zener references as transfer standards, shows good agreement to about 0.05 ppm between NML (ROC) and the other national laboratories
  • Keywords
    Josephson effect; measurement standards; voltage measurement; 1 V; 1991; China; ETL; Germany; Japan; Josephson-array voltage standard; NIST; NML; NPL; National Measurement Laboratory; PTB; UK; USA; Zener references; international comparisons; national laboratories; transfer standards; Atomic clocks; Hardware; Helium; Laboratories; Measurement standards; Microwave oscillators; NIST; Tail; Thermal conductivity; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278554
  • Filename
    278554