• DocumentCode
    1059604
  • Title

    A calibration method for four-terminal-pair high-frequency resistance standards

  • Author

    Suzuki, Kiyoshi ; Aoki, Toshiaki ; Yokoi, Katsumi

  • Author_Institution
    Yokogawa-Hewlett-Packard Ltd., Tokyo, Japan
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    379
  • Lastpage
    384
  • Abstract
    A calibration method for four-terminal-pair (4TP) resistance standards (4TP-R-STDs) is developed for frequencies up to about 10 MHz. The calibration is performed using an error-correction equation for a 4TP system. This is derived from analysis of an eight-port error adapter. The derivation of the uncertainties is also discussed
  • Keywords
    calibration; electric resistance measurement; measurement errors; measurement standards; 10 MHz; 4TP-R-STDs; calibration; eight-port error adapter; error-correction equation; four-terminal-pair high-frequency resistance standards; uncertainties; Admittance; Bridge circuits; Calibration; Capacitance; Equations; Error correction; Frequency; Impedance; Measurement standards; Standards development;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278586
  • Filename
    278586