DocumentCode
1059604
Title
A calibration method for four-terminal-pair high-frequency resistance standards
Author
Suzuki, Kiyoshi ; Aoki, Toshiaki ; Yokoi, Katsumi
Author_Institution
Yokogawa-Hewlett-Packard Ltd., Tokyo, Japan
Volume
42
Issue
2
fYear
1993
fDate
4/1/1993 12:00:00 AM
Firstpage
379
Lastpage
384
Abstract
A calibration method for four-terminal-pair (4TP) resistance standards (4TP-R-STDs) is developed for frequencies up to about 10 MHz. The calibration is performed using an error-correction equation for a 4TP system. This is derived from analysis of an eight-port error adapter. The derivation of the uncertainties is also discussed
Keywords
calibration; electric resistance measurement; measurement errors; measurement standards; 10 MHz; 4TP-R-STDs; calibration; eight-port error adapter; error-correction equation; four-terminal-pair high-frequency resistance standards; uncertainties; Admittance; Bridge circuits; Calibration; Capacitance; Equations; Error correction; Frequency; Impedance; Measurement standards; Standards development;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.278586
Filename
278586
Link To Document