DocumentCode :
1059604
Title :
A calibration method for four-terminal-pair high-frequency resistance standards
Author :
Suzuki, Kiyoshi ; Aoki, Toshiaki ; Yokoi, Katsumi
Author_Institution :
Yokogawa-Hewlett-Packard Ltd., Tokyo, Japan
Volume :
42
Issue :
2
fYear :
1993
fDate :
4/1/1993 12:00:00 AM
Firstpage :
379
Lastpage :
384
Abstract :
A calibration method for four-terminal-pair (4TP) resistance standards (4TP-R-STDs) is developed for frequencies up to about 10 MHz. The calibration is performed using an error-correction equation for a 4TP system. This is derived from analysis of an eight-port error adapter. The derivation of the uncertainties is also discussed
Keywords :
calibration; electric resistance measurement; measurement errors; measurement standards; 10 MHz; 4TP-R-STDs; calibration; eight-port error adapter; error-correction equation; four-terminal-pair high-frequency resistance standards; uncertainties; Admittance; Bridge circuits; Calibration; Capacitance; Equations; Error correction; Frequency; Impedance; Measurement standards; Standards development;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.278586
Filename :
278586
Link To Document :
بازگشت